Myron-l 3P User Manual

Browse online or download User Manual for Tools Myron-l 3P. Myron L 3P User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 13
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Ultrameter
Operation
Manual
Model 3P
10 - 02 (WEB) EG
MYRON L
COMPANY
Page view 0
1 2 3 4 5 6 ... 12 13

Summary of Contents

Page 1 - Model 3P

Ultrameter™ Operation ManualModel 3P10 - 02 (WEB) EGMYRON LCOMPANY

Page 2 - ULTRAMETER

2. Pour DI water in the sensor well and allow to stand for no more than 4 hours (longer can deplete the reference solution and damage the glass bulb

Page 3 - ULTRAMETER MODELS

XII. ACCESSORIESA. pH Buffer SolutionspH buffers are available in pH values of 4, 7 and 10. Myron L Companybuffer solutions are traceable to NIS

Page 4

the lifetime of a pH junction from depletion of solution inside thereference junction or from contamination. The junction is damage

Page 5

Like acidity and alkalinity, the increase of one is at the expense of theother, so a single voltage is called the Oxidation-Reduction Potent

Page 6

MSMRpHMMYYRROONN LL CCOOMMPPAANNYY3PORPThese 2 Measurement keys will:• Turn instrument on• Measure parameter• Exit any functionParameters (2)Wris

Page 7

FEATURES and SPECIFICATIONSA. Features• Accuracy of ±.01 pH/±1 mV @ calibration point.• All electrodes are internal for maximum protection.• Wa

Page 8

TABLE OF CONTENTSInstrument Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1FEATURES and SPECIFI

Page 9 - Figure 8

I. INTRODUCTIONThank you for selecting the Ultrameter™ Model 3P, one of the Myron LCompany’s latest in a new line of instruments utilizing

Page 10 - Top View

2. Rinse sensor well 3 times with sample to be measured. Shake out each sample to remove any residual liquid.3. Refill sensor well with sample.4. Pres

Page 11

6. Press to accept the new value. The pH Zero Calibration is now complete. You may continue with pH Gain Calibration or exit by pressing ei

Page 12 - Figure 9

Single point Gain Calibration is complete. You may continue for the 3rd point of Calibration (2nd Gain) or exit by pressing either measurement key. Ex

Page 13

VII. CHANGING from CENTIGRADE to FAHRENHEIT1. Press .2. Press to display the stored memory records.3. Press r

Comments to this Manuals

No comments