RECOMMENDED STANDARD SOLUTIONS
MODEL/RANGE SOLUTION NUMBER
EP
x .1 None available
x 1 None available
x 10 442-30
x 100 442-300
x 1000 442-3000
with RE-10 442-30,000
EP-10
x 1 None available
x 10 KCl-70
x 100 KCl-700
x 1000 KCl-7000
with RE-10 KCl-70,000
STANDARD SOLUTION EQUIVALENT VALUES
Solution Micromhos
Number Microsiemens PPM/NaCl PPM/442™*
442-30 46.7 21.8 30
442-300 445 214 300
442-3000 3900 2027 3000
442-30,000 30,100 18,235 30,000
KCl-70 70 32.8 45
KCl-700 700 340 478
KCl-7000 7000 3740 5687
KCl-70,000 70,000 47,999 84,983
* Myron L Company proprietary Natural Water Standard
RANGE EXTENDER
The RE-10 Range Extender is a useful accessory for
testing high conductivity solutions beyond the normal
range of your EP meter. Inserting the RE-10 into a sample-
lled cell cup increases the maximum range ten times.
Use the Range Extender whenever the reading is off the
scale.
USING THE RE-10
1. Fill the cell cup three times to rinse it, but each time
insert the Range Extender to rinse it also.
2. Fill the cell cup with your sample. Push the Range
Extender into the cell cup, seating the O-ring seal.
3. Use and read the meter in the normal manner. Multiply
the reading by 10.
4. For best accuracy, repeat the complete test with a fresh
sample.
5. When you’re done testing, remove the Range Extender.
Thoroughly rinse the cell cup and Extender with clean
water (preferably distilled or deionized) to eliminate
dried salts build-up. This is extremely important when
the instrument will be used to test high purity water.
CALIBRATING THE RE-10
1. Calibrate your meter (without RE-10) using Standard
Solution number 442-3000 (EP) or KCL-7000 (EP10).
2. Fill the cell cup with 442-30,000 high conductivity
Standard Solution.
3. Insert RE-10 and press the button. Multiply the reading
by 10 and compare it with the value on the Standard
Solution label. It they are not the same the RE-10 must
be recalibrated.
NOTE: DO NOT adjust the Calibration Control in your
meter.
4. Adjust the white insert of the Extender as follows:
If the reading is too high — push or tap inward.
If the reading is too low — twist or pull outward with
pliers.
NOTE: Each Range Extender is calibrated to a particular
meter. It should be recalibrated if it is to be used with
another Myron L meter.
PORTA PAK
Carrying Case for use with all
Myron L portable meters, is foam-
lined and molded of sturdy ABS
plastic.
ORDERING
To order accessories or instruments contact your
nearest stocking distributor, or the Myron L Company.
IF YOU ALSO TEST pH
Our reliable Model EP11/pH pDS meter may be ideal for
your application.
Same four conductivity
ranges as EP10 plus
2-12 pH
pH sensor built-in for
maximum protection,
but is eld replaceable
Automatic temperature
compensation
RE-10 RANGE EXTENDER
ACCESSORIES CONTINUED
2450 Impala Drive
Carlsbad, CA 92010-7226
TELEPHONE: +1-760-438-2021
FAX: +1-760-931-9189
www.myronl.com
© Myron L Company 2008
EPOM WEB 05-08
Specications subject to change without noticePrinted in the U.S.A.
Page view 3
1 2 3 4

Comments to this Manuals

No comments