8. WhentheLEDashesslowly,swirlthepenaroundtoremovebubbles,keepingthe
cellsubmerged.TheCalibrationscreenwilldisplay.“Calibrating 7.0 pH ...”
9. WhentheLEDlightgoessolidON,theCalibrationscreenwilldisplay
“7.0 ph CAL SAVED”alongwiththebeforeandaftercalibrationvalues.
10. DONEandCONTINUEButtonswillappearatthebottomoftheCalibrationScreen.
NOTE:TaptheDONEbuttontoendthecalibrationatthispoint,or;
11. Rinse the pH sensor withthe next pH buffer (it can be either 4.0 pH or 10.0 pH
buffer).
12. TaptheCONTINUEButtontocalibratethenextpoint.
13. Repeatsteps7trough10.Theinstrumentwill
automaticallyrecognizethecalibrationpointbasedonthebufferused.
14. Afterthe3rdcalibrationpointissuccessfullysavedonlythe
DONEbuttonwillappear.Tappingthisbuttonendsthe
calibration.
C. Single-Point Calibration.
1. Usingeither4,7,or10pHbuffer,performthecalibrationasdescribedinsteps1
through9ofSectionB.
2. Afterthecalibrationpointissuccessfullysaved,taptheDONEbutton
D. Factory Calibration
WhenpHbuffersarenotavailable,thePTBT2canbereturnedtofactorydefault
calibrationusingtheFACCALfunction.Thiswillerasestoredwetcalibrationconstants.
NOTE:defaultfactorycalibrationresetstheelectronicsonlyanddoesNOTtakethe
conditionofthesensorintoconsideration.
1. InsidetheApptaptheCALbuttonintheFeatureNavigationBar.TheAppscreen
willsay,“Paired ULTRAPEN is Ofine.”
2. Pressandreleasethepushbutton.Thescreenwilldisplay,“Press and hold the
button on the ULTRAPEN to enter CAL mode”.
3. PressandHoldtheULTRAPEN’sbuttontheLEDwillgotosolidONandtheApp’s
Calibrationscreenwillappear.
4. TaptheredFAC CALbutton.Awarningdialoguebox
willappear.
5. TaptheOKbuttoninthedialoguebox.TheApp’sCalibratingscreenwilldisplay,
“FAC CAL Reset”.
MAINTENANCE
I. Battery Replacement
ThePTBT2App displayhas anindicator thatdepicts the penbattery’s chargelevel.
Whenthechargelevelfallsbelow25%,theindicatorwillashred.Imediatelyreplace
thebatterywithannewNtypebattery.
1. InaCLEAN DRYplaceunscrewthebatterycapinacounter-clockwisemotion.
2. Slidethebatteryandbatteryhousingoutofthepen.
3. Removethedepletedbatteryfromitshousing.
4. Insertanewbatteryintothebatteryhousingorientedwiththenegativeendtouching
thespring.
5. Alignthegroovealongthebatteryhousingwiththeguidebumpinsidethepencase
andslidethebatteryhousingbackin.
6. Screwthebatterycapbackoninaclockwisedirection.
... Continued from page 3
II. Routine Maintenance
1. ALWAYSrinsethepHsensorwithcleanwateraftereachuse.
2. ALWAYSreplacethesoakercaphalffilledwithpHSensorStorageSolutionto
preventthesensorfromdryingoutaftereachuse.
3. Ifthesensorbecomesdirty,cleanthesensorsurfacewithanisopropylsoaked
cottonswab.Thenrinsethoroughlywithcleanwater.
4. Donotdrop,throworotherwisestrikethepen.Thisvoidsthewarranty.
5. Donotstorethepeninalocationwheretheambienttemperaturesexceedits
specifiedOperating/StorageTemperaturelimits.
Do not over tighten.