II. FEATURES and SPECIFICATIONS
A. Features
Ranges:
Conductivity/TDS — 0- 20,000 µS/ppm (TPH1 & TP1)
pH — 0 -14 (TPH1 & TH1)
Superior resolution four (4) digit LCD
Conductivity/TDS accuracy of
±
1% of reading
pH accuracy of
±
.01 pH units (TPH1 & TH1)
All electrodes are internal for maximum protection
Waterproof to 1 meter/3 feet
Memory storage (20 readings)
Autoranging Conductivity/TDS (TPH1 & TP1)
Easy keypad calibration
Prompts for simple pH calibration (TPH1 & TH1)
Factory calibrations stored in microprocessor
User selectable Conductivity/TDS modes (TPH1 & TP1)
3 “User Selectable” solution conversions (tempcos) (TPH1 & TP1)
User Selectable “Salinity” units (TPH1 & TP1)
Temperature accuracy of
±
0.1°C/F
Automatic Temperature Compensation to 25°C
Temperature Compensation disable feature
B. General Specications
Display 4 Digit LCD
Dimensions (LxW xH) 196 x 68 x 64 mm
7.7 x 2.7 x 2.5 in.
Weight 320 g/11.2 oz.
Case Material ABS
Cond/TDS Cell Material ABS
Cond/TDS Electrodes 316 Stainless Steel
Cond/TDS Cell Capacity 5 ml/0.2 oz.
pH Sensor Well Capacity 1,2 ml/0.04 oz. (TPH1 & TH1)
Power 9V Alkaline Battery
Battery Life >100 Hours/5000 Readings
Operating/Storage Temperature 0-55°C/ 32-132°F
Protection Ratings IP67/NEMA 6
(waterproof to 1 meter/3 feet)