4. Press ; the display will change to the other unit.
5. Press
; all temperature reading are now in degrees last
shown.
NOTE: Tempco will still be shown in %/°C
.
XII. TOTAL RETURN to FACTORY SETTINGS “FAC SEL”
There may come a time when it would be desirable to quickly reset
all the recorded calibration values in the instrument back to the factory
settings. This might be to ensure all calibrations are set to a known
value, or to give the instrument to someone else free of adjustments or
recorded data for a particular application.
NOTE: All stored data will be lost.
1. Press
.
2. Press to display the stored memory records.
3. Press repeatedly until
you pass the CLr ALL
and the C-F locations. The
display will show a
FAC SEL” (see Figure 19).
4. Press
to accept the resetting. Display will return to
Conductivity mode.
XIII. CELL CHECK
The cell check veries the cleanliness of the conductivity/TDS/MIN/SALT
sensor. In normal use the cell may become dirty or coated and require
cleaning. If the display is showing .00when the cell cup is dry, the sensor
is probably clean. No matter what a manufacturer claims, a sensor can
and will become contaminated or coated; therefore require cleaning. A
true 4-wire sensor, as in the Po o l Pr o , helps to mitigate contamination,
however, NO SENSOR IS 100% IMMUNE.